Membrane based on decyl-18-crown-6 for a potassium selective sensor

Wybrane pełne teksty z tego czasopisma
Identyfikatory
Warianty tytułu
PL
Membrana stosująca jako jonofor decylo-18-korona-6 jako selektywny czujnik potasowy
Języki publikacji
EN
Abstrakty
EN
Four kinds of 18-crown-6 derivatives were studied as ionophores in plasticized poly(vi-nyl chloride) (PVC) in order to obtain potassium selective membrane. The properties of the sensor were studied in detail. For the optimization of the selectivity toward K+ over Na+ the influence of the following factors was studied: nature and concentration of the ionophore; nature and concentration of the plasticizer and lipophilic salt addition. The potassium ion-selective membrane electrode was designed. The electrode has a fast and Nernstian response in the 1 x 10(-1)-2 x 10(-5) mol 1(-1) K+, after six months of continuous measurements. The change of pH in the range 2-10.5 does not affect the response of the electrode. The internal resistance of the electrode was measured. The electrode shows a good stability and can be used in the potentiometric determination.
PL
Badano cztery rodzaje pochodnych związku: 18-korona-6 jako jonofory w plastyfi-kowanym polichlorku winylu (PVC), w celu uzyskania membrany selektywnej względem jonów potasowych. Szczegółowo badano własności czujnika pomiarowego. W celu optymalizacji selektywności względem jonów K+ w porównaniu do jonów Na+ zbadano wpływ następujących parametrów: rodzaj i stężenie jonoforu, rodzaj i stężenie plasty-fikatora, oraz dodatek soli lipofilowej. Zaprojektowano jonoselektywną elektrodę membranową, selektywną względem jonów potasowych. Elektroda ma krótki czas odpowiedzi i nernstowską odpowiedź w granicach stężeń: 10(-1) x 10(-5) mol l(-1) K+ po sześciu miesiącach ciągłych pomiarów. Zmiana pH w granicach 2-10.5 nie wpływa na odpowiedź elektrody. Zmierzono wewnętrzną rezystancję elektrody. Elektrodę cechuje dobra trwałość i można ją używać do oznaczeń potencjometrycznych.
Czasopismo
Rocznik
Strony
41--49
Opis fizyczny
Bibliogr. 16 poz.
Twórcy
  • Institute of Chemistry, Fântânele 30, 3400 Cluj-Napoca, România
autor
  • Technical University of Cluj-Napoca, Muncii 105, 3400 Cluj-Napoca, România
autor
  • Institute of Chemistry, Fântânele 30, 3400 Cluj-Napoca, România
autor
  • Institute of Chemistry, Fântânele 30, 3400 Cluj-Napoca, România
  • Institute of Chemistry, Fântânele 30, 3400 Cluj-Napoca, România
Bibliografia
  • 1. Umezawa Y., Ion-selective Electrodes: Selectivity Coefficients, CRC Press, Florida 1990, p. 578.
  • 2. Cattrall R.W., Tribuzio S. and Freiser H., Anal. Chem., 46, 2223 (1974).
  • 3. Tamura H., Kimura K. and Shano T., Bull. Chem. Soc. Jap., 53, 547 (1980).
  • 4. Bratov A., Abramova N., Munoz J., Dominguez C. Alegret S. and Bartroli J., Anal. Chem., 67, 3589 (1995).
  • 5. Ammann D., Morf W.E., Anker P., MerIer P.C., Pretsch E. and Simon W., Ion-selective Electrode Rev., 5, 3 (1983).
  • 6. Oggenfuss P., Morf W.E., Oesch U., Ammann D., Pretsch E. and Simon W., Anal. Chim. Acta, 180, 299 (1986).
  • 7. Rechnitz G.A. and Eyal E., Anal. Chem., 44, 370 (1972).
  • 8. Oesch U., Ammann D. and Simon W., Clin. Chem., 32, 1448 (1986).
  • 9. Arnold M.A. and Meyerhoff M.E., Anal. Chem., 56, 20R (1984).
  • 10. Attyat A.S., Chrstian G.D., Cason C.V. and Bartsch R.A., Electroanalysis, 4, 51 (1992).
  • 11. Pioda L.A.R. and Simon W., Chemia, 24, 72 (1979).
  • 12. Hassan, Saad S.M., Mahmoud Wagiha H. and Othman Abdel Hameed M., Talanta, 44, 1087 (1997).
  • 13. IUPAC, Analytical Chemistry Division, Commission on Analytical Nomenclature, Pure Appl. Chem., 67, 507 (1995).
  • 14. Coşofreţ V.V., Nahir T.M., Lindner E. and Buck R.P., J. Electroanal. Chem., 327, 137 (1992).
  • 15. Oesch U. and Simon W., Anal. Chem., 52, 692 (1980).
  • 16. Ammann D., Pretsch E., Simon W., Lindner E., Bezegh A. and Pungor E., Anal. Chim. Acta, 171, 119 (1985).
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-article-BPP1-0022-0039
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.