Improving the clinical physiological examination environment - Avoiding EMD on medical equipment used for detecting biological signals

Wybrane pełne teksty z tego czasopisma
Identyfikatory
Warianty tytułu
PL
Poprawa środowiska elektromagnetycznego klinicznych badań fizjologicznych – Unikanie wpływu zaburzeń elektromagnetycznych na urządzenia medyczne do detekcji sygnałów biologicznych
Języki publikacji
EN
Abstrakty
EN
Shimane University Hospital constructed an electromagnetic shielded room to do electromyogram examinations. However, the quality of physiological examinations to detect weak biosignals was not acceptable. To create a safer environment, we re-installed the electric grounding and shortened the power cable to the minimum. And, we covered the power cable using electromagnetic shielding material. Also, we used an electromagnetic shielded mesh tent with shielding capacity of 20-30 dB. As a result, the electromagnetic noise was much reduced.
PL
Szpital Uniwersytecki Shimane zbudował ekranowane elektromagnetycznie pomieszczenie do badań elektromiograficznych. Jednak jakość badań fizjologicznych podczas detekcji słabych biosygnałów nie była zadowalająca. W celu poprawy środowiska badań zainstalowano uziom, skrócono i zaekranowano przewody zasilające oraz zastosowano namiot z siatki ekranującej o skuteczności ekranowania 20-30 dB. W rezultacie znacznie zredukowano szum elektromagnetyczny.
Rocznik
Strony
64--66
Opis fizyczny
Bibliogr. 7 poz., rys.
Twórcy
autor
autor
autor
autor
Bibliografia
  • [1] Merletti R., Botter A., Troiano A., et al. Technology and instrumentation for detection of the surface electromyographic signal: State of the art. Clinical Biomechanics 24, 2009, pp. 122-134.
  • [2] Clancy E.A., Morin E.L., Merletti R. Sampling, noise-reduction and amplitude estimation issues in surface electromyography. Journal of Electromyography and Kinesiology 12, 2002, pp.1-16.
  • [3] MauguieÁre F., Allison T., Babilon C. Somatosensory evoked potentials. Recommendations for the Practice of Clinical Neurophysiology: Guidelines of the IFCN, Chapter 2.4, Elsevia Science, 1999, pp.79-90.
  • [4] Vincent A. Methods for improving the signal-to-noise ratio of endogenous-evoked potentials. Integrative Physiological and Behavioral Science 27(1), 1992, pp.54-65
  • [5] Nuwer MR. Aminoff M., Desmedt J. et al, IFCN recommended standards for short latency somatosensory evoked potentials. Report of an IFCN committee. Electroencephalography and Clinical Neurophysiology, 1994, 91(1), pp.6-11.
  • [6] Hanada E. The electromagnetic environment of hospitals: how it is affected by the strength of electromagnetic fields generated both inside and outside the hospital. Ann Ist Super Sanità 2007, 43(3), pp.208-217
  • [7] Hanada E., Takano K., Antoku Y., et al. Electromagnetic shielding capacity of stainless-steel fiber mesh. IEEE EMC Society Newsletter, 2003, 196, pp.21-26
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-article-BPOK-0039-0014
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.