Investigation of pitting and crevice corrosion resistance of NiTi alloy by means of electrochemical methods

Autorzy
Wybrane pełne teksty z tego czasopisma
Identyfikatory
Warianty tytułu
PL
Badania odporności stopu NiTi na korozję wżerową oraz szczelinową z wykorzystaniem metod elektrochemicznych
Języki publikacji
EN
Abstrakty
EN
Corrosion resistance of implant alloys is a very important determinant of their biocompatibility. The nature of the environment and surface treatments have a significant influence on corrosion. The main aim of the research was evaluation of pitting and crevice corrosion resistance of NiTi alloy in simulated physiological solution. The evaluation of the electrochemical behavior of NiTi alloy was realized by recording of anodic polarization curves with the use of the potentiodynamic and chronomaperometric methods.
PL
Odporność korozyjna stopów implantacyjnych stanowi podstawowy czynnik determinujący ich biokompatybilność. Zasadniczy wpływ na odporność korozyjną posiadają charakter środowiska pracy oraz zastosowane zabiegi obróbki powierzchniowej implantów. Celem przeprowadzonych badań było określenie odporności stopu NiTi o zmodyfikowanych powierzchniach na korozję wżerową oraz szczelinową z wykorzystaniem metody potencjodynamicznej oraz chronoamperometrycznej.
Rocznik
Strony
102--105
Opis fizyczny
Bibliogr. 29 poz., rys., tab.
Twórcy
autor
Bibliografia
  • [1] Gil F.J., Planell J.A., Shape memory alloys for medical applications. Proc Instn Mech Engrs, Vol 212 Part H, 473 – 488
  • [2] Shabalovskaya S.A., Surface, corrosion and biocompatibility aspects of Nitinol as an implant material. Bio-Medical Materialsand Engineering, 12 (2002), 69–109
  • [3] Haarsters J., Salis-Solio G., Bensmann G., The use of NiTi as an implant material in orthopaedics. In ShapeMemory in Engineering Aspects of Shape Memory Alloys (Eds T. W. Duering, K. N. Melton, D. Stockel and C. M. Wayman), (1990), 426–427 (Butterworth-Heinemann, London)
  • [4] Sekiguci Y., Medical Applications in Shape Memory Alloys (Ed. H. Funakubo), (1984), 10–23 (Gordon and Breach Science Publishers, London), 142–149
  • [5] Hughes J.L., Evaluation of Nitinol for use as a material in the construction of Orthopaedic Implants, DAMD 74-C-4041 US Army Medical Research and Development Command, Fort Detrick, Frederick, Maryland, (1977), 72–78
  • [6] Haasters J., Baumgart F., Bensmann G., Memory alloys—new material for implantation in orthopedic surgery, Part 2. In Current Concepts of Internal Fixation of Fractures (Ed. H. K. Uthoff ), (1980), 128–130 (Springer-Verlag, New York).
  • [7] Schmerling M.A., Wilkov M.A., Sandres, A.E., Wooseley, J.E., Using the shape recovery of Nitinol in the Harrington rod treatment of scoliosis. J. Biomed. Mater. Res., (1976), 10, 879–902
  • [8] Sawyer P.N., Page M., Rudewald B., Lagergren H., Baselius L., McCool C., Halperin W., Srinivasan S., Characteristics of the human heart: design requirements for replacement. Trans. Am. Soc. Artif. Int. Organs, (1971), 17, 470–471
  • [9] Netsu N., Iwabuchi T., Honma, T., A study of the NiTi intracranial aneurysm clip having the shape memory effect. Bull. Res. Inst. Min. Dress. Met. Tohotu Univ., (1978), 34, 67–68
  • [10] Fukuyo S., Suzuki Y., Suzuki K. Sairenji E., Shape memory implants. In Shape Memory in Engineering Aspects of Shape Memory Alloys (Eds T. W. Duering, K. N. Melton, D. Sto¨ ckel and C. M.Wayman), (1990), 470–472 (Butterworth-Heinemann, London)
  • [11] Gil F.J., Fernandez E., Manero J.M., Planell J.A., Sabria J., Cortada M., Giner L., A study of load cyclic in a NiTi alloy with pseudoelastic behaviour used in dental prosthetic fixators. Biomed. Mater. Engng, (1996), 6(3), 153–158
  • [12] Miura F., Mogi M., Ohura Y., Karibe M., The superelastic japaneese NiTi alloy wire for use in orthodontics. Am. J. Orthod. Dent. Orthop., (1988), 94(2), 89–96
  • [13] Libenson C., Gil F.J., Planell J.A., Untersuchung der Nickel- Titandraehte auf ihre Verwendbarkeit in der Klinischen Kieferorthopaedie. Orthod. Kieferorthop., (1996), 28(1), 135–154
  • [14] Kajzer W., Kaczmarek M., Krauze A., Marciniak J., Surface modification and corrosion resistance of Ni-Ti alloy used for urological stents. Journal of Achievements in Materials and Manufacturing Engineering, Vol. 20, 1-2, (2007), 123-126
  • [15] Kaczmarek M., Walke W., Kajzer W., Chemical composition of passive layers formed on metallic biomaterials. Archives of Material Science and Engineering, Vol. 20, 5, (2007), 273-276 PRZEGLĄD ELEKTROTECHNICZNY (Electrical Review), ISSN 0033-2097, R. 86 NR 12/2010 105
  • [16] M.Kaczmarek, Corrosion resistance of NiTi alloy in simulated body fluids. Archives of Materials Science and Engineering, Vol. 20, 5, (2007), 269-272
  • [17] Kaczmarek M., Problematyka doboru własności stopu NiTi do zastosowań w chirurgii rekonstrukcyjnej. Inżynieria Biomateriałów, (2007), 65-66, 55-56
  • [18] Rondelli G., Vicentini B., (1999) Biomaterials 20:785–792
  • [19] Kim H., Johnson W.J., (1999) Angle Orthod 69(1):39–44
  • [20] Es-Souni M., Es-Souni M., Fischer-Brandies H., (2002) Biomaterials, 2887–2894
  • [21] Es-Souni M., Fischer-Brandies H., Es-Souni M. (2003) J Orofac Orthop 64,16–25
  • [22] Wever D.J., Veldhuizen A.G., de Vries J., Busscher H.J., Uges D.R.A., van Horn J.R., (1998) Biomaterials,19,761–769
  • [23] Trepanier C., Tabrizian M., Yahia L.H., Bilodou L., Piron D.L., (1998) J Biomed Mater Res, 43, 433–440
  • [24] Speck K.M., Fraker A.C., (1980) J Dent Res, 59(10),1590–1595
  • [25] Rondelli G., (1996) Biomaterials 17, 2003–2008
  • [26] Venugopalan R., Trepanier C., (2000) Min Invas Ther Allied Technol 9(2), 67–74
  • [27] Kuphasuk C., Ossida Y., Andrei C.J., Hovijitra S.T., Barco M.T., Brown D.T., J Prosthet Dent, (2001), 85(2),195–202
  • [28] El Medawar L., Rocher P., Hornez J.C., Traisnel M., Breme J., Hildebrand H.F., Biomol Eng, (2002) 19,153–160
  • [29] Montero-Ocampo C., Lopez H., Salinas Rodriguez A., J Biomed Mater Res, (1996), 32,583–591
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-article-BPOK-0032-0028
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.