PL EN


Preferencje help
Widoczny [Schowaj] Abstrakt
Liczba wyników
Tytuł artykułu

Electrical parameters measurement and testing of implantable cardioverters – defibrillators

Wybrane pełne teksty z tego czasopisma
Identyfikatory
Warianty tytułu
PL
Pomiar parametrów elektrycznych i testowanie implantowanych cardio-stymulatorów
Języki publikacji
EN
Abstrakty
EN
The aim of the work was measurement and testing of implantable cardioverter - defibrillator (ICDs) device. There was realised electronic measuring circuit which substitutes human body tissue during the device accuracy testing. This circuit enables to measure defibrillation shock discharge in the whole energy range (0.1- 41J). The presented measuring and testing procedures can easy check-up precision of parameters and functionality of embedded electronic device circuit’s part of the ICDs. Presented suggestion and results of this work can be implemented as verification procedure in various types of ICDs and also pacemakers before their implantation as well.
PL
Celem pracy było zbadanie właściwości elektrycznych rozrusznika serca. Zaproponowano procedure testowania tego typu urzadzeń.
Rocznik
Strony
186--189
Opis fizyczny
Bibliogr. 7 poz., rys., tab., wykr.
Twórcy
autor
autor
autor
  • VSB - Technical Unviversity Ostrava, Faculty of Electrical Engineering and Computer Science, Department of Measurement and Control,17. listopadu 15, 70833 Ostrava - Poruba, Marek.Penhaker@vsb.cz
Bibliografia
  • [1] Korpas, D., Psychological Intolerance to Implantable Cardioverter Defibrillator, In Biomedical Papers- Olomouc, Volume 152, Issue: 1 p.: 147-149 , June 2008 , ISSN: 1213-8118
  • [2] Palan, B., Roubik, K., Husak, M., Courtois, B., CMOS ISFETbased structures for biomedical applications, In Proceedings Medicine & Biology at 1st Annual International IEEE-Embs Special Topic Conference on Microtechnologies, Pages: 502- 506 Published: 2000, ISBN: 0-7803-6603-4
  • [3] Bax JJ, Abraham T, Barold SS, Breithardt OA, Fung JW, Garrigue S, Gorcsan J 3rd, Hayes DL, Kass DA, Knuuti J, Leclercq C, Linde C, Mark DB, Monaghan MJ, Nihoyannopoulos P, Schalij MJ, Stellbrink C, Yu CM. Cardiac resynchronization therapy: Part 1--issues before device implantation. J Am Coll Cardiol. 2005 Dec 20;46(12):2153-67.
  • [4] Augustynek, M., Penhaker, M., Korpas, D. Controlling Peacemakers by Accelerometers. In 2010 The 2nd International Conference on Telecom Technology and Applications, ICTTA 2010. March 19-21, 2010, Bali Island, Indonesia, Volume2 NJ. IEEE Conference Publishing Services, 2010, p. 161–163. ISBN 978-0-7695-3982-9, DOI: 10.1109/ICCEA.2010.288
  • [5] Ector H, Rickards AF, Kappenberger L, et al. The world survey of cardiac pacing and implantable cardioverter defibrillators: calendar year 1997 – Europe. Pacing Clin Electrophysiol 2001;24(5):863–8.
  • [6] Guidant Corporation Cardiac Rhythm Management Technical Services. Impact of Therapeutic Radiation and Guidant ICD/CRTD/CRT-P/Pacing systems. St. Paul, MN, Guidant Corporation, 2004, Revision; pp. 1–6.
  • [7] Labza, Z., Penhaker, M., Augustynek, M., Korpas, D. Verification of Set Up Dual-Chamber Pacemaker Electrical Parameters. In 2010 The 2nd International Conference on Telecom Technology and Applications, ICTTA 2010. March 19- 21, 2010, Bali Island, Indonesia, Volume 2NJ. IEEE Conference Publishing Services 2010, p. 168–172. ISBN 978-0-7695-3982-9., DOI: 10.1109/ICCEA.2010.187
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-article-BPOC-0058-0032
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.