The paper reports experimental results, which demonstrate harmful influence of a single NdFeB magnet to some measuring devices. The NdFeB magnet used in the experiment (N42, D50 mm x h30 mm) is of average size available commercially. The experimental results allow concluding that a single NdFeB magnet is able to permanently damage an energy meter (Ferraris and with current transformer based on Co core) and low-voltage current transformer with a silicon-iron core. In case of the examined Ferraris meter, the most susceptible places appear to be the braking magnet and magnetic circuits of the current and voltage coils.
PL
W pracy przedstawiono wyniki badań obrazujące szkodliwy wpływ pojedynczego magnesu ziem rzadkich NdFeB na różnego rodzaju urządzenia pomiarowe. Wykorzystany podczas badań magnes NdFeB (N42, D50 mm x h30 mm) jest powszechnie dostępnym magnesem o średniej gęstości energii. Zaprezentowane wyniki badań pozwalają na wysunięcie wniosku, że pojedynczy magnes NdFeB może spowodować uszkodzenie licznika energii elektrycznej (zarówno indukcyjnego, jak i z przekładnikiem prądowym o rdzeniu Co), jak również przekładnika prądowego niskiego napięcia z rdzeniem SiFe. W przypadku liczników indukcyjnych najbardziej podatnymi na oddziaływanie pola magnetycznego są: magnes hamujący oraz obwody magnetyczne cewek prądowej I napięciowej.
Rocznik
Strony
88--92
Opis fizyczny
Bibliogr. 15 poz., rys., tab., wykr.
Twórcy
autor
autor
autor
autor
autor
  • Czestochowa University of Technology
Bibliografia
  • [1] Seminar, Protection direction of Watt-hour meters from influence of external magnetic field, Polish Chapter of the IEEE Magnetics Society, Czestochowa, 20.05.2005.
  • [2] M. Soinski and W. Pluta, Amorphous Co-based core under influence of DC magnetic field, in Proc. 17th Soft Magnetic Materials Conference, 07 - 09.09.2005 Bratislava, pp. 216
  • [3] F. Garbalinski and S. Warzecha, EMC defect in watt-hour meter (Ferraris) operating in strong stationary magnetic field, in Proc. 18th International Wroclaw Symposium and Exhibition on Electromagnetic Compatibility, Wroclaw, 28-30 June 2006.
  • [4] W. Pluta, R. Rygal, M.Soinski and S. Zurek, Shielding effect of current transformers used in electronic electrical energy meters from impact of external NdFeB magnets, in Proc. 18th International Wroclaw Symposium and Exhibition on Electromagnetic Compatibility, Wroclaw, 28-30 June 2006.
  • [5] P. Kepski, R. Pytlech, W. Pluta, R. Rygal, M. Soinski and S. Zurek, Hazardous influence of rare-earth magnets on low-voltage current transformer as the evidence of EMC compatibility lack, in Proc. 18th International Wroclaw Symposium and Exhibition on Electromagnetic Compatibility, Wroclaw, 28-30 June 2006.
  • [6] M. Soinski, R. Rygal, W. Pluta, P. Kepski and S. Zurek Magnetic toroidal cores exposed to a strong stationary magnetic field, Magnetic Measurements, Závaná Poruba, Slovakia, September 21 - 23, 2006.
  • [7] Rare-earth Permanent Magnets VACODYM, VACOMAX, Catalogue PD-002, Vacuumschmelze GmbH&Co. KG, 2003.
  • [8] M. Soinski and A. J. Moses, Anisotropy of Iron-based Soft Magnetic Materials, Chapter 4, Handbook of Magnetic Materials, Vol. 8, North-Holland Elsevier, 1995.
  • [9] Industrial Research Project no. BZ-3-306-3/2005, Department of Electrical Engineering, Technical University of Czestochowa, 2006
  • [10] Current transformers for electronic Watt-hour meters, Catalogue PB-CT, Vacuumschmelze GmbH&Co. KG, 2002.
  • [11] D. Jiles, Introduction to Magnetism and Magnetic Materials, Chapman & Hall, London (1991).
  • [12] Manual to Flux2D Program, CEDRAT, France (1998).
  • [13] IEC 62053-21, Electricity metering equipment (a.c.) - Particular requirements - Part 21: Static meters for active energy (class 1 and 2), 2003.
  • [14] R. Pytlech and R. Rygal and W. Pluta and M. Soinski, Quick measurement procedure for quality flux density checking at 0,3 A/m in toroidal cores, Journal of Electrical Engineering (Chech and Slovak Centre of IEE), Vol. 55, No 10/5, 2004.
  • [15] P. Kepski, The comparison of current transformer abilities under influence of rare-earth magnet, Seminar, Technical University of Czestochowa, 30.01.2006.
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-article-BPOC-0022-0007
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.