Round window membrane motion before and after stapedotomy surgery - an experimental study

Wybrane pełne teksty z tego czasopisma
Identyfikatory
Warianty tytułu
Języki publikacji
EN
Abstrakty
EN
This paper presents the results of non-contact measurements of the human ear round window (RW) membrane motion in freshly harvested human cadaver temporal bone specimens. A PSV 400 Scanning Laser Vibrometer system (Polytec, Waldbronn, Germany) was used to determine the effect of stapedotomy on the sound-induced displacement amplitude of the RW membrane. The vibration patterns were measured in the frequency range of 0.4–10 kHz in four specimens with air conduction stimulation at 90 dB SPL applied to the external auditory canal. It was shown that the vibration amplitude of the RW membrane after stapes piston prosthesis implantation, in comparison with a physiological specimen, was reduced several times. The motion of the RW membrane can be used to determine the cochlear stimulation for the evaluation of middle ear ossicle reconstruction, especially in the case of otosclerosis surgery.
Rocznik
Strony
27--33
Opis fizyczny
Bibliogr. 14 poz., il.
Twórcy
autor
autor
autor
  • Institute of Micromechanics and Photonics, Faculty of Mechatronics, Warsaw Technical University, Poland, m.kwacz@mchtr.pw.edu.pl
Bibliografia
  • [1] PURIA S., PEAKE W., ROSOWSKI J., Sound-pressure measurements in the cochlear vestibule of human-cadaver ears, J. Acoust. Soc. Am., 1997, 101 (5), 2754-2770.
  • [2] SLAMA M., Middle ear pressure gain and cochlear input impedance in the chinchilla, Thesis, Massachusetts Institute of Technology, 2008.
  • [3] AIBARA R., WELSH J., PURIA S., GOODE R., Human middleear sound transfer function and cochlear input impedance, Hearing Research, 2001, 152 (1-2), 100-109.
  • [4] MERCHANT S., RAVICZ M., ROSOWSKI J., Acoustic input impedance of the stapes and cochlea in human temporal bones, Hearing Research, 1996, 97 (1-2), 30-45.
  • [5] KRINGLEBOTN M., The equality of volume displacement in the inner ear windows, J. Acoust. Soc. Am., 1995, 98, 192-196.
  • [6] STENFELT S., HATO N., GOODE R., Fluid volume displacement at the oval and round windows with air and bone conduction stimulation, J. Acoust. Soc. Am., 2004, 115, 797-812.
  • [7] STENFELT S., HATO N., GOODE R., Round window membrane motion with air conduction and bone conduction stimulation, Hearing Research, 2004, 198, 10-24.
  • [8] ROSOWSKI J., DAVIS P., MERCHANT S., DONAHUE K., COLTRERA M., Cadaver middle ears as models for living ears: comparison of middle ear input immitance, Ann. Otol. Rhinol. Laryngol., 1990, 99, 403-412.
  • [9] GOODE R., BALL G., NISHIHARA S., Measurement of umbo vibration in human subjects - method and possible clinical applications, Am. J. Otol., 1993, 14, 247-251.
  • [10] GOODE R., BALL G., NISHIHARA S., NAKAMURA K., Laser Doppler Vibrometer (LDV) - A new clinical tool for the otologist, Otol. Neurotol., 1996, 17 (6), 813-822.
  • [11] NISHIHARA S., GOODE R., Measurement of tympanic membrane vibration in 99 human ears, [in:] Research and Otosurgery: Proceedings of the International Workshop on Middle Ear Mechanics in Research and Otosurgery, K.B. Huttenbrink (editor), Dresden University Press, Dresden, Germany, 1997, 91-93.
  • [12] HUBER A., LINDER T., FERRAZZINI M., SCHMID S., DILLIER N., FISCH U., Intraoperative scanning laser Doppler interferometry, [in:] Recent Developments in Auditory Mechanics, H. Wada, T. Takasaka, K. Ikeda, K. Ohyama, T. Koike (editors), World Scientific, Singapore, 2000, 10-14.
  • [13] KHANNA S., TONNDORF J., The vibratory pattern of the round window in cats, J. Acoust. Soc. Am., 1971, 50, 1475-1483.
  • [14] NOMURA Y., Otological significance of the round window, Adv. Oto. Rhino. Laryngol., 1984, 33, 1-162.