PL EN


Preferencje help
Widoczny [Schowaj] Abstrakt
Liczba wyników
Tytuł artykułu

Photoelastic investigation of mechanical problems in applying acetabulum cups to hip prosthesis

Wybrane pełne teksty z tego czasopisma
Identyfikatory
Warianty tytułu
Języki publikacji
EN
Abstrakty
EN
Acetabulum cups are the counterparts of endoprostheses, which are mounted in the pelvis. There are different designs of such cups. They differ in their form and in the way of mounting. This paper deals with the threaded acetabulum cups with conical shape mounted in the pelvis. In general, one has to distinguish between stresses caused by the mounting process and lifetime stresses of the mounted cup. Stresses remaining of the mounting must ensure the so-called primary stability; however, they will disappear to some degree due to rheological properties of bone. The question arises whether the mounting stresses are responsible for local plastic deformations and the corresponding residual stresses or not. In this paper, mostly loading stresses and their dependence on the contact properties between pelvis and cup or between cup and polyurethane inlay are considered.
Rocznik
Strony
47--54
Opis fizyczny
Bibliogr. 8 poz., rys., wykr.
Twórcy
autor
autor
autor
  • Laboratory of Experimental Mechanics, University of Zagreb, Croatia
Bibliografia
  • [1] LERF R., Prüfbericht ID 8256, Metallographie Proj. No. PL 04, PRECICION IMPLANT AH, Aarau/Switzerland, 29.06.2004.
  • [2] ROHR L., Prüfbericht der Eidgenössischen Materialprüfungs- und Forschungsanstalt EMPA, Thun/Switzerland, 2004.
  • [3] BEER R.J., JECIC S., KODVANJ J., GRIENAUER W., Application of partially polymerised Araldite B in photoelasticity by means of ionising radiation (Primjena djelomi_no polimeriziranog araldita B u radijacijskoj fotoelasticimetriji), 1999, Strojarstvo, Vol. 41, No. 5/6, pp. 173–179.
  • [4] JECIC S., BEER R.J., KODVANJ J., FLEISCHMANN M., Shrinking of polymers in technical problems consisting of two different parts determined by 3D photoelasticity using gamma irradiation, Proceedings of the 17th IMEKO World Congress, Dubrovnik, 2003, pp. 1935–1938.
  • [5] KLANKE J., HÖRMANSDÖRFER G., WESTERMANN K., Biomechanische Prüfung von Schraubpfannen an geeigneten Knochenersatzmaterialien, Schraubpfannen, State of the Art, Editor: Effenberger H., MCU, 2004, pp. 95–102.
  • [6] WITZEL U., RIEGER W., EFFENBERGER H., Dreidimensionale Spannungsanalyse bei Schraubpfannen und ihren knöchernen Lagern, Schraubpfannen, State of the Art, Editor: Effenberger H., MCU, pp.77-86, 2004.
  • [7] RHO J.Y., KUHN-SPEARING L., ZIOUPOS P., Mechanical properties and the hierarchical structure of bone, Med. Engineering & Physics, 1998, 20, pp. 92–102.
  • [8] FROCHT M.M., Photoelasticity, Vol. II, John Wiley & Sons Inc., New York, 1949.
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-article-BPB1-0028-0045
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.