PL EN


Preferencje help
Widoczny [Schowaj] Abstrakt
Liczba wyników
Tytuł artykułu

Helmholtz coils system for testing field meters used to measure magnetic fields generated by traction vehicles

Treść / Zawartość
Identyfikatory
Warianty tytułu
PL
Układ cewek Helmholtza do testowania mierników pola stosowanych do pomiaru pól magnetycznych generowanych przez pojazdy trakcyjne
Języki publikacji
EN
Abstrakty
EN
This paper presents a Helmholtz coils system for testing field meters used to measure magnetic fields generated by traction vehicles. The results of conducted studies of the levels of magnetic fields generated by electrical and electronic devices in the railway environment indicate the need for control tests of systems for measuring electromagnetic fields. The test stand developed for controlling magnetic field tests ensures the correctness, reliability, and repeatability of these tests.
PL
W artykule przedstawiono układ cewek Helmholtza zastosowany do testowania mierników pola stosowanych do pomiaru pól magnetycznych generowanych przez pojazdy trakcyjne. Wyniki przeprowadzonych badań poziomów pól magnetycznych generowanych przez urządzenia elektryczne i elektroniczne w środowisku kolejowym wskazują na konieczność przeprowadzania badań kontrolnych układów do pomiarów pól elektromagnetycznych. Opracowane stanowisko do kontroli pomiarów pola magnetycznego zapewnia poprawność, wiarygodność i powtarzalność tych badań.
Rocznik
Strony
147--151
Opis fizyczny
Bibliogr. 15 poz., rys., tab.
Twórcy
  • Sieć Badawcza Łukasiewicz - Poznański Instytut Technologiczny ul. Ewarysta Estkowskiego 6, 61-755 Poznań
  • Politechnika Poznańska, Instytut Elektrotechniki i Elektroniki Przemysłowej, ul. Piotrowo 3a, 60-965 Poznań
Bibliografia
  • [1] PN-EN 50500:2008//A1:2015-10 Measurement procedures of magnetic field levels generated by electronic and electrical apparatus in the railway environment with respect to human exposure
  • [2] Yang Y., Song Z., Jiang L., Rao B. and Zhang M., An Improved Two-Coil Configuration for Low-Frequency Magnetic Field Immunity Tests and Its Field Inhomogeneity Analysis, IEEE Trans. Industr. Electr., 65 (2018), No. 10, 8204-8214
  • [3] Beiranvand R., Effects of the Winding Cross-Section Shape on the Magnetic Field Uniformity of the High Field Circular Helmholtz Coil Systems, IEEE Trans. Industr. Electr., 64 (2017), No. 9, 7120-7131
  • [4] Ishii M., Suzuki M., Impedance of Helmholtz Coil to Generate Standard AC Magnetic Field in High Frequency, Conference on Precision Electromagnetic Measurements (CPEM), USA, 2020, 1-2
  • [5] Fabricio de Melo C., Araújo R. L., Morozowski L., Nilton S., Quoirin R., Ikeda M., Costa A., A., Calibration of low frequency magnetic field meters using a Helmholtz coil, Measurement, 42 (2009), No. 9, 1330- 1334
  • [6] Yang Li, Jun Xu, Xiangyu Kang, Zhengkun Fan, Xiangmei Dong, Xiumin Gao, Songlin Zhuang, Design of highly uniform three-dimensional square magnetic field coils for external magnetic shielding of magnetometers, Sensors and Actuators A: Physical, 331 (2021),113037
  • [7] Haghnegahdar, A., Khosrovpanah, H., Andisheh-Tadbir, A., Mortazavi, G., Saeedi Moghadam, M., Mortazavi, S., Zamani, A., Haghani, M., Shojaei Fard, M., Parsaei, H., & Koohi, O., Design and fabrication of helmholtz coils to study the effects of pulsed electromagnetic fields on the healing process in periodontitis: preliminary animal results. Journal of biomedical physics & engineering, 4 (2014), nr 3, 83–90
  • [8] Kossowski T., Szczupak P., Analysis of the influence of strong magnetic field on unmanned aircrafts, using Helmholtz coil Przegląd Elektrotechniczny, 96 (2020), nr 6, 11-14
  • [9] ISO/IEC 17025:2017 General requirements for the competence of testing and calibration laboratories
  • [10] Garczarek A., Stachowiak D., Badanie i analiza pól magnetycznych generowanych przez pojazdy trakcyjne w odniesieniu do narażenia ludzi, Przegląd Elektrotechniczny, 98 (2022), nr 8, 179-184
  • [11] Cvetkovic D., Cosic I., Modelling and Design of Extremely Low Frequency Uniform Magnetic Field Exposure Apparatus for In Vivo Bioelectromagnetic Studies, Engineering in Medicine and Biology Society, Conference of the IEEE, 2007, 1675-1678
  • [12] Bansal R., Ed., Handbook of Engineering Electromagnetics. Dekker Publ., New York, 2004
  • [13] ICNIRP Guidelines for limiting exposure to time-varying electric and magnetic fields (1 Hz – 100 kHz), Health Physics, 99(6) (2010), 818-836
  • [14] Directive 2013/35/EU of the European Parliament and of the Council of 26 June 2013 on the minimum health and safety requirements regarding the exposure of workers to the risks arising from physical agents (electromagnetic fields) (20th individual Directive within the meaning of Article 16(1) of Directive 89/391/EEC) and repealing Directive 2004/40/EC
  • [15] Dz.U. 2018 r. poz. 1286 Rozporządzenie Ministra Rodziny, Pracy i Polityki Społecznej z dnia 12 czerwca 2018 r. zmieniające rozporządzenie w sprawie najwyższych dopuszczalnych stężeń i natężeń czynników szkodliwych dla zdrowia w środowisku pracy
Uwagi
Opracowanie rekordu ze środków MNiSW, umowa nr POPUL/SP/0154/2024/02 w ramach programu "Społeczna odpowiedzialność nauki II" - moduł: Popularyzacja nauki i promocja sportu (2025).
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-8deec8ca-6851-41ae-bd64-d0ed904e7d78
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.