Endoprosthesis stem fractures are among the rarest complications that occur after hip joint arthroplasty. The aim of this paper is to evaluate the causes of the fractures of the Aura II stem neck, which is an element of an endoprosthesis implanted in a patient. In order to achieve it, a radiogram was evaluated, the FEM analysis was carried out for the hip joint replaced using the Aura II prosthesis and scanning tests as well as a chemical analysis were performed for the focus of fatigue. The tests performed indicate that the most probable causes leading to the fatigue fracture of the Aura II stem under examination were material defects in the process of casting and forging (forging the material with delamination and the presence of brittle oxides and carbides) that resulted in a significant reduction of strength and resistance to corrosion. In the light of an unprecedented stem neck fracture, this information should be an indication for non-destructive tests of ready-made stems aiming to discover the material and technological defects that may arise in the process of casting and drop forging.
Rocznik
Tom
Strony
223--234
Opis fizyczny
Bibliogr. 30 poz., rys., tab., wykr.
Twórcy
autor
- AGH University of Science and Technology Faculty of Mechanical Engineering and Robotics 30 Mickiewicz Ave., 30-059 Cracow, POLAND
autor
- AGH University of Science and Technology Faculty of Mechanical Engineering and Robotics 30 Mickiewicz Ave., 30-059 Cracow, POLAND
autor
- Cracow University of Technology, Faculty of Mechanical Laboratory of Coordinate Metrology 37 John Paul II Ave., 31-864 Cracow, POLAND
autor
- AGH University of Science and Technology Faculty of Non-Ferrous Metals 30 Mickiewicz Ave., 30-059 Cracow, POLAND
autor
- Jagiellonian University Medical College Faculty of Medicine, Dental Institute, Department of Dental Prosthodontics ul. Montelupich 4, 31-155 Cracow, POLAND
Bibliografia
- [1] NFZ - Statistics of JGP 2009 and 2010. http://prog.nfz.gov.pl/app-jgp/.2011.
- [2] Affatato S., Ruggiero A. and Merola M. (2015): Advanced biomaterials in hip joint arthroplasty. A review on polymer and ceramics composites as alternative bearings. - Composites Part B: Engineering, vol.83, pp.276-283.
- [3] Phedy P., Ismail H.D., Hoo C. and Djaja Y.P. (2017): Total hip replacement: A meta-analysis to evaluate survival of cemented, cementless and hybrid implants. - World Journal of Orthopedics, vol.8, No.2, pp.192.
- [4] Ollivere B., Darrah C., Barker T., Nolan J. and Porteous M.J. (2009): Early clinical failure of the Birmingham metal-on-metal hip resurfacing is associated with metallosis and soft-tissue necrosis. - Journal of Bone & Joint Surgery, British Volume, vol.91, No.8, pp.1025-1030.
- [5] Klein G.R., Levine B.R., Hozack W.J., Strauss E.J., D'Antonio J.A., Macaulay W. and Di Cesare P.E. (2007): Return to athletic activity after total hip arthroplasty: consensus guidelines based on a survey of the Hip Society and American Association of Hip and Knee Surgeons. - The Journal of Arthroplasty, vol.22, No.2, pp.171-175.
- [6] Chatterji U., Ashworth M.J., Lewis P.L. and Dobson P.J. (2005): Effect of total knee arthroplasty on recreational and sporting activity. - ANZ Journal of Surgery, vol.75, No.6, pp.405-408.
- [7] Söderman P., Malchau H., Herberts P. and Johnell O. (2000): Are the findings in the Swedish National Total Hip Arthroplasty Register valid?: A comparison between the Swedish National Total Hip Arthroplasty Register, the National Discharge Register, and the National Death Register. - The Journal of Arthroplasty, vol.15, No.7, pp.884-889.
- [8] Malchau H., Herberts P., Eisler T., Garellick G. and Söderman P. (2002): The Swedish total hip replacement register. - J. Bone Joint Surg. Am, vol.84, No.2, pp.S2-S20.
- [9] Hailer N.P., Garellick G. and Kärrholm J. (2010): Uncemented and cemented primary total hip arthroplasty in the Swedish Hip Arthroplasty Register: evaluation of 170,413 operations. - Acta Orthopaedica, vol.81, No.1, pp.34-41.
- [10] Suckel A., Geiger F., Kinzl L., Wulker N. and Garbrecht M. (2009): Long-term results for the uncemented Zweymuller/Alloclassic hip endoprosthesis: a 15-year minimum follow-up of 320 hip operations. - The Journal of Arthroplasty. vol.24, No.6, pp.846-853.
- [11] Kotela A., Ambroziak P. and Deszczyński M.J. (2012): Fracture of the hip joint endoprosthesis - case report - Sharp Duty, vol.5, pp.1-2.
- [12] Yates P.J., Quraishi N.A., Kop A., Howie D.W., Marx C. and Swarts E. (2008): Fractures of modern high nitrogen stainless steel cemented stems: cause, mechanism, and avoidance in 14 cases. - The Journal of Arthroplasty, vol.23, No.2, pp.188-196.
- [13] Jarvi K. and Kerry R.M. (2007): Segmental stem fracture of a cemented femoral prosthesis. - The Journal of Arthroplasty. vol.22, No.4, pp.612-616.
- [14] Harvie P., Haroon M., Henderson N. and El-Guindi M. (2007): Fracture of the hydroxyapatite-ceramic-coated JRI-Furlong femoral component. - Bone & Joint Journal, vol.89, No.6, pp.742-745.
- [15] Wroblewski B.M., Siney P.D. and Fleming P.A. (2016): Fracture of the stem. In Charnley Low-Frictional Torque Arthroplasty of the Hip. - Springer International Publishing, pp.187-201.
- [16] Busch C.A., Charles M.N., Haydon C.M., Bourne R.B., Rorabeck C.H., Macdonald S.J. and McCalden R.W. (2005): Fractures of distally-fixed femoral stems after revision arthroplasty. - Bone & Joint Journal, vol.87, No.10, pp.1333-1336.
- [17] Efe T. and Schmitt J. (2011): Analyses of prosthesis stem failures in noncemented modular hip revision prostheses. - The Journal of Arthroplasty, vol.26, No.4, pp.665-e7.
- [18] Drobniewski M., Sibiński M., Plebański R. and Synder M. (2010): Fracture of the endoprosthesis as a rare complication of the hip arthroplasty - a description of two cases. - Chir Narz Ruchu Ortop Pol, vol.75, pp.53-56.
- [19] Ryniewicz A.M., Madej T., Ryniewicz A. and Bojko Ł. (2016): A Biometrological Procedure Preceeding the Resurfacing. - Metrology and Measurement Systems, vol.23, No.1, pp.97-106.
- [20] Ryniewicz A., Ryniewicz A.M., Madej T., Sładek J. and Gąska A. (2013): Biometrological method of pelvis measurement and anatomical positioning of endoprosthesis of hip joint. - Metrology and Measurement Systems, vol.20, No.1, pp.17-26.
- [21] Korkmaz O., Bozkus F.S., Aykut U.S., Caki H.C. and Kaygusuz M.A. (2012): The effect of component position on clinical outcomes in resurfacing hip arthroplasty. - Acta Orthopaedica et Traumatologica Turcica, vol.46, No.1, pp.13-16.
- [22] http://www.biometorthopaedics.ch/resource/16323/pricelist.pdf.
- [23] www.stainless.eu/assets/files/produits/en/stainless-bars-M30NW.pdf.
- [24] Ryniewicz A.M. and Madej T. (2017): FEM analysis in the hip joint reconstructed with classical and modified hip resurfacing. - Engineering Transactions, vol.65, No.1, pp.123-131.
- [25] Ryniewicz A.M. (2011): Identification, modelling and biotribology of human joints. - AGH University of Science and Technology Press, Kraków.
- [26] Cilingir A.C. (2010): Finite element analysis of the contact mechanics of ceramic-on-ceramic hip resurfacing prostheses. - J Bionic Eng, vol.7, No.3, pp.244-53.
- [27] Walicka A. (2012): Inertia effects in porous squeeze film biobearing with rough surfaces lubricated by a powerlaw fluid. - Special Topics & Reviews in Porous Media: An International Journal, vol.3, No.3, pp.247-256.
- [28] Cwanek J. (2009): The usability of the surface geometry parameters for the evaluation of the artificial hip joint wear. - Rzeszów University Press.
- [29] Szala G. and Ligaj B. (2014): Models of fatigue life curves in fatigue life calculations of machine elements - examples of research. - Science notebooks. Transport / Silesian University of Technology, vol.82, pp.251-259.
- [30] Norman P., Iyengar S., Svensson I. and Flivik G. (2014): Fatigue fracture in dual modular revision total hip arthroplasty stems: failure analysis and computed tomography diagnostics in two cases. - The Journal of Arthroplasty, vol.29, No.4, pp.850-855.
Uwagi
PL
Opracowanie rekordu w ramach umowy 509/P-DUN/2018 ze środków MNiSW przeznaczonych na działalność upowszechniającą naukę (2018)
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-70d6316a-eb1b-4871-8944-d78bdfaf418d