PL EN


Preferencje help
Widoczny [Schowaj] Abstrakt
Liczba wyników
Tytuł artykułu

Biomechanical properties of hip implant with ceramics coating

Identyfikatory
Warianty tytułu
PL
Właściwości biomechaniczne przeszczepu kości biodrowej z powłoką ceramiczną
Języki publikacji
EN
Abstrakty
EN
In this paper the results of investigations of biomechanical properties of transplantation of hip bone with ceramic coating are present ed. The finite element analysis of the stress-strain state of the femur bone after hip replacement surgery and full recovery period are carried out. A finite element model of the femur bone is obtained on the basis of tomographic data of 36 years old male patient. A transplant stem with ceramic coating based on A400 lateralized specification was analyzed. Stresses in the intact femur and femur after arthroplasty were determined. The effect of the reduction of bone density as the result of removal of normal stresses by an implant were analyzed. The study serves as a biomechanical basis for development of artificial prostheses and for clinical hip joint replacements.
PL
W artykule przedstawiono wyniki badań właściwości biomechanicznych przeszczepu kości biodrowej z powłoką ceramiczną. Przeprowadzono analizę metodą elementów skończonych stanu naprężeń i odkształceń kości udowej po zabiegu wymiany stawu biodrowego i po pełnym okresie rekonwalescencji. Model kości udowej wykorzystany do analizy metodą elementów skończonych otrzymano na podstawie danych termograficznych 36-letniego pacjenta. Analizie poddano rdzeń przeszczepu okryty powłoką ceramiczną na podstawie specyfikacji A400. Wyznaczono naprężenia w kości udowej nienaruszonej i po endoprotezoplastyce. Analizowano wpływ zmniejszenia gęstości kości jako wyniku usunięcia naprężeń normalnych przez implant. Badania stanowią podstawę biomechaniczną do opracowywania sztucznych protez stawu biodrowego i klinicznych wymian połączenia biodrowego.
Rocznik
Strony
65--72
Opis fizyczny
Bibliogr. 18 poz., rys., tab., wykr.
Twórcy
  • Rzeszow University of Technology, 8 Powstańców Warszawy Av., 35-959 Rzeszów
autor
  • Rzeszow University of Technology
Bibliografia
  • [1] Ridzwan M.I.Z., Shuib S., Hassan A.Y., Shokri A.A., Mohamad Ibrahim M.N.: Problem of stress shielding and improvement to the hip implant designs: A Review – J. Medical Sci., 7 (2007) 460-467.
  • [2] Herrera A., Mateo J, Gil-Albarova J., Lobo-Escolar A., Ibarz E., Gabarre S., Más Y., Gracia L.: Cementless Hydroxyapatite Coated Hip Prostheses, BioMed Research Int., 2015 (2015) Article ID 386461, 13 pages.
  • [3] Nikitsin A.V., Mikhasev G.I., Maslov A.P: Finite element analysis of the porous coating in hip-joint prosthesis, Int. Sci. Tech. J. "Mechanics of Machines, Mechanisms and Materials", 18 (2012) 86-89.
  • [4] Mohammed R.A.: Interface micromotion of cementless hip stems in simulated hip arthroplasty, American J. Appl. Sci., 6 (2009) 1682-1689.
  • [5] Tonino A.J., Therin M., Doyle: Hydroxyapatite-coated femoral stems. Histology and histomorphometry around five components retrieved at post mortem, J. Bone Joint Surgery, 81 (1999) 148-154.
  • [6] Alexander R.: Animals mechanics, Singwick & Jackson, London 1968.
  • [7] Fonseca E.M.M., Lima M.J., Barreirra L.M.S.: Human femur assessment using isotropic and orthotropic materials dependent of bone density, 3rd Int, Conf, on Integrity, Reliability and Failure, Porto 2009, pp.7-8.
  • [8] Orlovskii V.P., Komlev V.S., Barinov S.M.: Hydroxyapatite and hydroxyapatite-based ceramics, Inorganic Materials, 38 (2002) 973-984.
  • [9] Huiskes R., Rietbergen B.: Preclinical testing of total hip stems. The effects of coating placement, Clinical Orthopaedics Related Res., 319 (1995) 64-76.
  • [10] Rosenthall L., Bobyn J.D., Tanzer M.: Bone densitometry: influence of prosthetic design and hydroxyapatitecoating on regional adaptive bone remodelling, Int. Orthopaedics, 23 (1999) 325-329.
  • [11] Nakashima Y., Sato T., Yamamoto T. : Results at aminimum of 10 years of follow-up for AMS and PerFix HA-coated cementless total hip arthroplasty: impact of cross-linkedpolyethylene on implant longevity, J. Orthopaedic Sci., 18 (2013) 962-968.
  • [12] Park Y.S., Lee J.Y., Yun S.H., Jung M.W., Oh I.: Comparison of hydroxyapatite- and porous-coated stems in total hip replacement, Acta Orthopaedica Scandinavica, 74 (2003) 259-263.
  • [13] Tas A.C., Ozgur E.: Manufacture of macroporous calcium hydroxyapatite bioceramics, J. Eur. Ceram. Soc., 19 (1999) 2569-2772.
  • [14] Yosibash Z.: A CT-based high-order finite element analysis of the human proximal femur compared to in vitro experiments, J. Biomech., 129 (2007), 297-309.
  • [15] Bergmann G.: Hip contact forces and gait patterns from routine activities, J. Biomech., 34 (2001) 859-871.
  • [16] Bojescul J.A.: Results of porous-coated anatomic total hip arthroplasty without cement at fifteen years: a concise follow-up of a previous report, The J. Bone and Joint Surgery, 85 (2003) 1079-1083.
  • [17] Duchemin L.: Prediction of mechanical properties of cortical bone by quantitative computed tomography, Medical Eng. Physics, 30 (2008) 21-328.
  • [18] Simões J.A., Vaz M.A., Blatcher S., Taylor M.: Influence of head constrain and muscle forces on the strain distribution within the intact femur, Medical Eng. Physics, 22 (2000) 453-459.
Uwagi
PL
Opracowanie ze środków MNiSW w ramach umowy 812/P-DUN/2016 na działalność upowszechniającą naukę (zadania 2017).
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-52084497-d465-42a4-9c1e-534c2c867eff
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.