Resorbable Tightening of Blood Vessel Protheses Prepared from Synthetic Polymers

Identyfikatory
Warianty tytułu
PL
Opracowanie resorbowalnego uszczelnienia protez naczyniowych z polimerów syntetycznych
Języki publikacji
EN
Abstrakty
EN
A method was prepared for the sealing of artificial blood vessels by combined soaking and spraying. Biocompatible and resorbable copolymers of poly(DL-lactide-co-glycolide) with varying contents of two co-monomers were used in the investigations. The materials were provided by Boehringer Ingelheim Co, Germany under the trade name Resomer. The polyester double-sided veloured knitted fabric vessel prosthesis DALLON H, by TRICOMED Co, constituted the matrix. Application trials resulted in the selection of Resomer RG 755S, containing 75% mol. residues of D,L-lactide and 25% mol. residues of glycolide. A modified polyester blood vessel prosthesis was prepared bearing the surgery tightness (water permeability below 30 ml/cm2/min) and physical-mechanical properties required, including elasticity adequate for such a medical device. Conditions for the sealing were elaborated at large laboratory scale with a notable composition of the coating bath and application of the coating. Results of the investigations can be considered as a basis for a technology layout for the sealing of artificial blood vessels with biocompatible and resorbable synthetic copolymer poly(DL-lactide-co-glycolide).
PL
Opracowano metodę uszczelnienia ścianek sztucznych naczyń krwionośnych przy wykorzystaniu resorbowalnego syntetycznego materiału polimerowego. Zastosowano biozgodne kopolimery poli(DL-laktydo-ko-glikolid) typu Resomer. Matrycę stanowiła poliestrowa dwustronnie welurowana dziana proteza naczyniowa DALLON H. Powłoka uszczelniająca była nanoszona na powierzchnię protezy kombinowaną techniką napawania i natrysku. W wyniku przeprowadzonych prac uzyskano modyfikowaną poliestrową protezę naczyniową, charakteryzującą się wymaganą szczelnością chirurgiczną (przepuszczalność wodna poniżej 30 ml/cm2/min) i odpowiednimi właściwościami fizyko-mechanicznymi. Wytypowane zostały warunki uszczelniania uwzględniające skład kąpieli powlekającej, metodykę nanoszenia powłoki uszczelniającej w skali wielkolaboratoryjnej. Przeprowadzone badania stanowią podstawę do przygotowania założeń technologii uszczelnienia sztucznych naczyń krwionośnych biozgodnym i resorbowalnym kopolimerem syntetycznym DL-laktyd/glikolid.
Rocznik
Strony
93--99
Opis fizyczny
Bibliogr. 26 poz., rys., tab.
Twórcy
  • Institute of Biopolymers and Chemical Fibres, Łódź, Poland.
autor
  • Institute of Biopolymers and Chemical Fibres, Łódź, Poland.
autor
  • Institute of Biopolymers and Chemical Fibres, Łódź, Poland.
  • Institute of Biopolymers and Chemical Fibres, Łódź, Poland.
Bibliografia
  • 1. Teng P., Feigin I,; J. Neurosurg 12, 1955 pp. 591–600.
  • 2. Harrison J. H.; AMA Arch Surg. 74(4), 1957 pp. 557-561.
  • 3. Taner D. Durmuş, Blood Vessel Substitutes http://www.biomed.metu.edu. tr/courses/term_papers/Blood-VesselSubstitutes_durmus.htm.
  • 4. Wesolowski S. A., Fries C. C., Karlson K. E., Bakey M., Sawyer P. N.; Surgery 50 (1961), pp. 91–96.
  • 5. Głowiński S., et al.: Hematologia, 14, 1981 pp. 393-398.
  • 6. Głowiński S.; Polimery w medycynie, 18, (4), 1988 pp. 221-235.
  • 7. M. Kucharska, Ph. D. Thesis“ Investigation in the use of chitosan for the modification of vessel prostheses made of polyester knitwear(in Polish), 1992 Technical University of Łódź.
  • 8. Ketharnathan V., Christie B. A.; Arch Surg 115 (1980) pp. 967–969.
  • 9. Woodhouse K.A., Klement P., Chen V., Gorbet M.B., Keeley F.W., et al.; BBiioommaa-terials, 25 (2004) pp. 4543–53.
  • 10. Patent USA 5, 108,424; 1996
  • 11. Patent USA 4, 747,848; 1988
  • 12. Patent USA 4, 902,29; 1990
  • 13. Patent WO 0117571; 2001
  • 14. Patent USA 5, 584,875,1996
  • 15. Greisler HP, et all.; Surgery 110(4), 1991 pp. 645-54.
  • 16. Yu TJ, et all.; J. Invest. Surg., 7(3), 1994 pp. 195-211.
  • 17. Perewgo G., et all; J. of Biomaterials Science, 14(10),2003 pp. 1057-1075.
  • 18. Lian Xue, Greisler H. P.; J. Vasc. Surg., 37 (2003) pp. 472-480.
  • 19. Greisler H.P., et all: J. of Biomed. Materials Research, 27(7), 2004 pp. 955-961.
  • 20. Middleton J. C., Tipton A. J.; Biomaterials 21 (2000), pp. 2335–2346.
  • 21. Lakshmi S. Nair, Cato T. Laurencin, Prog. Polym. Sci. 32 (2007) pp. 762-798.
  • 22. Wiliams D.F., Biomaterials Vol.: 30, Issue: 30, October, 2009, pp. 5897-5909.
  • 23. Conference proceedings 8th School of X-Ray Sterilization and Hygiene Promotion, Institute of Radiation Chemistry and Technique, 20-21.10.2005, Warsaw.
  • 24. Kroeze R. J., Helder M. N., Govaert L. E., Smit T. H.; Materials 2, (2009) pp. 833-856.
  • 25. Polish patent appl. 358157
  • 26. PN-79/P-04884.03 – Polish standard Methods of textile testing Knitted medical devices-prosthese of blood vessels Estimation of water permeability (in Polish).
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-34c8ba95-fa2f-44a9-97cf-0364da5839c6
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.