Synthetic Aperture Cardiac Imaging with Reduced Number of Acquisition Channels. A Feasibility Study

Treść / Zawartość
Identyfikatory
Warianty tytułu
Języki publikacji
EN
Abstrakty
EN
Commercially available cardiac scanners use 64-128 elements phased-array (PA) probes and classical delay-and-sum beamforming to reconstruct a sector B-mode image. For portable and hand-held scanners, which are the fastest growing market, channel count reduction can greatly decrease the total power and cost of devices. The introduction of ultra-fast imaging methods based on plane waves and diverging waves provides new insight into heart’s moving structures and enables the implementation of new myocardial assessment and advanced flow estimation methods, thanks to much higher frame rates. The goal of this study was to show the feasibility of reducing the channel count in the diverging wave synthetic aperture image reconstruction method for phased-arrays. The application of ultra-fast 32-channel subaperture imaging combined with spatial compounding allowed the frame rate of approximately 400 fps for 120 mm visualization to be achieved with image quality obtained on par with the classical 64-channel beamformer. Specifically, it was shown that the proposed method resulted in image quality metrics (lateral resolution, contrast and contrast-to-noise ratio), for a visualization depth not exceeding 50 mm, that were comparable with the classical PA beamforming. For larger visualization depths (80-100 mm) a slight degradation of the above parameters was observed. In conclusion, diverging wave phased-array imaging with reduced number of channels is a promising technology for low-cost, energy efficient hand-held cardiac scanners.
Rocznik
Strony
437--446
Opis fizyczny
Bibliogr. 14 poz., rys., wykr.
Twórcy
  • Institute of Fundamental Technological Research, Polish Academy of Sciences, Pawińskiego 5B, 02-106 Warsaw, Poland
  • Institute of Fundamental Technological Research, Polish Academy of Sciences, Pawińskiego 5B, 02-106 Warsaw, Poland
autor
  • Institute of Fundamental Technological Research, Polish Academy of Sciences, Pawińskiego 5B, 02-106 Warsaw, Poland
autor
  • Institute of Fundamental Technological Research, Polish Academy of Sciences, Pawińskiego 5B, 02-106 Warsaw, Poland
Bibliografia
  • 1. Bae M. H., Jeong M. K. (2000), A study of syntheticaperture imaging with virtual source elements in B-mode ultrasound imaging systems, IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 47, 6, 1510-1519.
  • 2. Cikes M., Tong L., Sutherland G. R., D’hooge J. (2014), Ultrafast Cardiac Ultrasound Imaging: Technical Principles, Applications, and Clinical Benefits, JACC: Cardiovascular Imaging, 7, 8, 812-823.
  • 3. Cygan S., Kumor M., Żmigrodzki J., Leśniak-Plewińska B., Kowalski M., Kałużyński K. (2017), Left ventricular phantoms with inclusions simulating transmural and non-transmural infarctions – FEM and EchoPAC study, Medical Imaging 2017: Ultrasonic Imaging and Tomography, 1013918-1.
  • 4. Hasegawa H., Kanai H. (2011), High-frame-rate echocardiography using diverging transmit beams and parallel receive beamforming, Journal of Medical Ultrasonics, 38, 3, 129-140.
  • 5. Lewandowski M., Walczak M., Witek B., Kulesza P., Sielewicz K. (2012), Modular & Scalable Ultrasound Platform with GPU Processing, Proc. 2012 IEEE Ultrasonics Symp., pp. 2071-2074.
  • 6. Moore C. et al. (2015), Live high-frame-rate echocardiography, IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 62, 10, 1779-1787.
  • 7. Papadacci C., Pernot M., Couade M., Fink M., Tanter M. (2014), High-contrast ultrafast imaging of the heart, IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 61, 2, 288-301.
  • 8. Phantom (2018), http://www.cirsinc.com/products/all/67/.
  • 9. Seraphim A., Paschou S. A., Grapsa J., Nihoyannopoulos P. (2016), Pocket-Sized Echocardiography Devices: One Stop Shop Service?, Journal of Cardiovascular Ultrasound, 24, 1, 1-6.
  • 10. Sulzbach-Hoke L. M., Schanne L. C. (1999), Using a portable ultrasound bladder scanner in the cardiac care unit, Critical Care Nurse, 6, 19, 35-39.
  • 11. Tasinkevych Y., Klimonda Z., Lewandowski M., Nowicki A., Lewin P. A. (2013), Modified multielement synthetic transmit aperture method for ultrasound imaging: A tissue phantom study, Ultrasonics, 53, 570-579.
  • 12. Tasinkevych Y., Trots I., Nowicki A., Lewandowski M. (2012a), Optimization of the Multi-element Synthetic Transmit Aperture Method for Medical Ultrasound Imaging Applications, Archives of Acoustics, 37, 1, 47-55.
  • 13. Tasinkevych Y., Trots I., Nowicki A., Lewin P. A. (2012b), Modified synthetic transmit aperture algorithm for ultrasound imaging, Ultrasonics, 52, 2, 333-342.
  • 14. Tong L., Gao H., Choi H. F., D’hooge J. (2012), Comparison of conventional parallel beamforming with plane wave and diverging wave imaging for cardiac applications: a simulation study, IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 59, 8, 1654-1663.
Uwagi
Opracowanie rekordu w ramach umowy 509/P-DUN/2018 ze środków MNiSW przeznaczonych na działalność upowszechniającą naukę (2018).
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-3029de26-a1bc-4700-801d-8257453541bb
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.