A method for acquiring a ballistocardiographic (BCG) signal from the feet of a standing person and the back of a sitting or lying patient is described. The measurements are carried out using in-house constructed fibre-optic sensors interrogated with a commercially available system. The sensor head consists of a fibre Bragg grating (FBG) attached to an elastic board that is placed between the monitored person's body and a soft surface, enabling the board to deform in an unobstructed way. The body's movements, including the BCG component, exert pressure on the board and make it deform along with the attached FBG. The changes to the Bragg wavelength are proportional to the body's movement and a BCG signal can be extracted from the obtained recording. The measuring capabilities of the sensors were evaluated by compar- ing the heart rate (obtained on the basis of the BCG signal) with the reference signal registered by an ECG recorder. An RMS value of the relative error is below 1.8% and statistical analyses show a satisfactory reconstruction of measurements. Tests carried out in the MRIenvironment proved the method to be immune to strong electromagnetic fields. The presence of the sensor in an MRI scanner does not affect the quality of imaging.
Twórcy
autor
  • Technical Department of Aeromedical Research and Flight Simulators, Military Institute of Aviation Medicine, ul. Zygmunta Krasińskiego 54/56, 01-755 Warszawa, Poland
  • Technical Department of Aeromedical Research and Flight Simulators, Military Institute of Aviation Medicine, Warsaw, Poland
Bibliografia
  • [1] Maniewski R. Magnetic examinations of electrical and mechanical cardiac activity. Wroclaw, Poland: Ossolineum; 1987 [monograph in Polish].
  • [2] Johnson CA, Laing GH. Simultaneous mechanograms and electrograms from the intact human subject, with notes on the effects of distention of the stomach on the conventional electrocardiogram: preliminary report. Am Heart J 1940;20:160–73.
  • [3] Goedhard WJ. Ballistocardiography: past, present and future. Bibl Cardiol 1979;37:27–45.
  • [4] Winer NJ. Direct body ballistocardiography: a 25 year survey. Landmarks in its representation of cardiac dynamics. Bibl Cardiol 1979;37:46–57.
  • [5] Winer NJ. High-frequency (acceleration) direct body ballistocardiography. Representation of the carotid pulse, phono- and apex-cardiogram in the BCG. Bibl Cardiol 1975;33:37–43.
  • [6] Bancroft WH. Kinetocardiography: past and present. Bibl Cardiol 1979;37:58–72.
  • [7] Juznic G. Genesis of the sphygmogram from the kinetocardiogram. Bibl Cardiol 1979;37:84–108.
  • [8] Giovangrandi L, Inan OT, Wiard RM, Etemadi M, Kovacs GTA. Ballistocardiography – a method worth revisiting. Proceedings of the 33rd Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Boston, USA, 30 Aug – 03 Sep; 2011. pp. 4279–82.
  • [9] Brüser C, Winter S, Leonhardt S. Robust inter-beat interval estimation in cardiac vibration signals. Physiol Meas 2013;34:123–38.
  • [10] Inan OT, Kovacs GTA. A low-noise ac-bridge amplifier for ballistocardiogram measurement on an electronic weighing scale. Physiol Meas 2010;31:N51–9.
  • [11] Eblen-Zajjur A. A simple ballistocardiographic system for a medical cardiovascular physiology course. Adv Physiol Educ 2003;27:224–9.
  • [12] Pandia K, Inan OT, Kovacs GTA, Giovangrandi L. Extracting respiratory information from seismocardiogram signals acquired on the chest using a miniature accelerometer. Physiol Meas 2012;33:1643–60.
  • [13] Chung GS, Lee JS, Hwang SH, Lim YK, Jeong D-U, Park KS. Wakefulness estimation only using ballistocardiogram: nonintrusive method for sleep monitoring. Proceedings of the 32nd Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Buenos Aires, Argentina, 31 Aug – 04 Sep; 2010. pp. 2459–62.
  • [14] Alametsä J, Varri A, Viik J, Hyttinen J, Palomäki A. Ballistocardiogaphic studies with acceleration and electromechanical film sensors. Med Eng Phys 2009;31:1154–65.
  • [15] Grillet A, Kinet D, Witt J, Schukar M, Krebber K, Pirotte F, et al. Optical fiber sensors embedded into medical textiles for healthcare monitoring. IEEE Sens J 2008;8:1215–22.
  • [16] Šprager S, Zazula D. Detection of heartbeat and respiration from optical interferometric signal by using wavelet transform. Comput Methods Programs Biomed 2013;111: 41–51.
  • [17] Ładyżyński P, Wójcicki JM, Chihara K. Application of telemedicine technique in screening for diabetic retinopathy. Biocybern Biomed Eng 2007;27:253–63.
  • [18] Chen Y, Yang H. Self-organized neural network for the quality control of 12-lead ECG signals. Physiol Meas 2012;33:1399–418.
  • [19] Dziuda L, Rozanowski K, Skibniewski FS, Krej M. The device for measuring vital signs. Patent application no. P 391763 submitted to the Patent Office of the Republic of Poland; 2010.
  • [20] Dziuda L, Skibniewski FW, Krej M, Lewandowski J. Monitoring respiration and cardiac activity using fiber Bragg grating based sensor. IEEE Trans Biomed Eng 2012;59:1934–42.
  • [21] Hill KO, Meltz G. Fiber Bragg grating technology fundamentals and overview. J Lightwave Technol 1997;15:1263–76.
  • [22] Majumder M, Gangopadhyay TK, Chakraborty AK, Dasgupta K, Bhattacharya DK. Fiber Bragg gratings in structural health monitoring – present status and applications. Sens Actuators A: Phys 2008;147:150–64.
  • [23] Fusiek G, Niewczas P, Judd MD. Towards the development of a downhole optical voltage sensor for monitoring electrical submersible pumps. Sens Actuators A: Phys 2012;184:173–81.
  • [24] Al-Fakih E, Abu Osman NA, Mahamd Adikan FR. The use of fiber Bragg grating sensors in biomechanics and rehabilitation applications: the state-of-the-art and ongoing research topics. Sensors 2012;12:12890–926.
  • [25] Fresvig T, Ludvigsen P, Steen H, Reikerås O. Fibre optic Bragg grating sensors: an alternative method to strain gauges for measuring deformation in bone. Med Eng Phys 2008;30:104–8.
  • [26] Mishraa V, Singh N, Rai DV, Tiwari U, Poddar GC, Jain SC, et al. Fiber Bragg grating sensor for monitoring bone decalcification. Orthop Traumatol Surg Res 2010;96:646–51.
  • [27] Betts RA, Tjugiarto T, Xue YL, Chu PL. Nonlinear refractive index in erbium doped optical fiber: theory and experiment. IEEE J Quantum Electron 1991;27:908–13.
  • [28] Xu MG. The measurement of physical fields using optical fibres and fibre gratings.[Ph.D. dissertation] Hampshire, UK: Dept Electron and Comput Sci, Univ Southampton; 1995.
  • [29] Moerman KM, Sprengers AMJ, Nederveen AJ, Simms CK. A novel MRI compatible soft tissue indentor and fibre Bragg grating force sensor. Med Eng Phys 2013;35:486–99.
  • [30] González-Landaeta R, Casas O, Pallàs-Areny R. Heart rate detection from an electronic weighing scale. Physiol Meas 2008;29:979–88.
  • [31] Inan OT, Etemadi M, Wiard RM, Giovangrandi L, Kovacs GTA. Robust ballistocardiogram acquisition for home monitoring. Physiol Meas 2009;30:169–85.
  • [32] Alametsä J, Viik J, Alakare J, Värri A, Palomäki A. Ballistocardiography in sitting and horizontal positions. Physiol Meas 2008;29:1071–87.
  • [33] Dziuda L, Skibniewski FW, Rozanowski K, Krej M, Lewandowski J. Fiber-optic sensor for monitoring respiration and cardiac activity. Proceedings of the IEEE Sensors Conference, Limerick, Ireland, 28 Oct – 31 Oct; 2011. pp. 413–6.
  • [34] Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;327:307–10.
  • [35] Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res 1999;8:135–60.
  • [36] Zazula D, Donlagic D, Sprager S. Application of fibre-optic interferometry to detection of human vital signs. J Laser Health Acad 2012;1:27–32.
  • [37] Prisk GK, Verhaeghe S, Padeken D, Hamacher H, Paiva M. Three-dimensional ballistocardiography and respiratory motion in sustained microgravity. Aviat Space Environ Med 2001;72:1067–74.
  • [38] Gubner RS, Rodstein M, Ungerleider HE. Ballistocardiography: an appraisal of technic, physiologic principles, and clinical value. Circulation 1953;7:268–86.
  • [39] March HW. Three-plane ballistocardiography: the effect of age on the longitudinal, lateral, and dorsoventral ballistocardiograms. Circulation 1955;12:869–82.
  • [40] Migeotte PF, Tank J, Pattyn N, Funtova I, Baevsky R, Neyt X, et al. Three dimensional ballistocardiography: methodology and results from microgravity and dry immersion. Proceedings of the 33rd Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Boston, USA, 30 Aug – 03 Sep; 2011. pp. 4271–4.
  • [41] Migeotte P-F, Deliere Quentin, Tank J, Funtova I, Baevsky R, Neyt X, et al. 3D-ballistocardiography in microgravity: comparison with ground based recordings. Proceedings of the 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Osaka, Japan, 3–7 July; 2013. pp. 7012–6.
  • [42] Dziuda Ł, Skibniewski FW, Krej M, Baran P. Fiber Bragg grating-based sensor for monitoring respiration and heart activity during magnetic resonance imaging examinations. J Biomed Opt 2013;18:057006.
  • [43] Harris LM, Robinson J, Menzies RG. Predictors of panic symptoms during magnetic resonance imaging scans. Int J Behav Med 2001;8:80–7.
  • [44] Meléndez JC, McCrank E. Anxiety-related reactions associated with magnetic resonance imaging examinations. J Am Med Assoc 1993;270:745–7.
  • [45] Meuret AE, Ritz T, Wilhelm FH, Roth WT. Voluntary hyperventilation in the treatment of panic disorder – functions of hyperventilation, their implications for breathing training, and recommendations for standardization. Clin Psychol Rev 2005;25:285–306.
  • [46] Pittig A, Arch JJ, Lam CW, Craske MG. Heart rate and heart rate variability in panic, social anxiety, obsessive– compulsive, and generalized anxiety disorders at baseline and in response to relaxation and hyperventilation. Int J Psychophysiol 2013;87:19–27.
  • [47] Shin JH, Hwang SH, Chang MH, Park KS. Heart rate variability analysis using a ballistocardiogram during Valsalva manoeuvre and post exercise. Physiol Meas 2011;32:1239–64.
  • [48] Silva AF, Carmo JP, Mendes PM, Correia JH. Simultaneous cardiac and respiratory frequency measurement based on a single fiber Bragg grating sensor. Meas Sci Technol 2011;22:075801.
  • [49] Zhang C, Miao CY, Li HQ, Song HC, Xu FJ. Smart textile sensing system for human respiration monitoring based on fiber Bragg grating. Proceedings of SPIE 2009;7381:738104.
  • [50] Hao J, Jayachandran M, Kng PL, Foo SF, Aung PW, Cai Z. FBG-based smart bed system for healthcare applications. Front Optoelectron China 2010;3:78–83.
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-2d948250-8425-41a3-9571-cdf08e03fe1e
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.