Tytuł artykułu

Dosimetric investigation of dual energy photon beams with assymmetric collimator jaws

Treść / Zawartość
Warianty tytułu
Języki publikacji
EN
Abstrakty
EN
Many modern linear accelerators are equipped with asymmetric collimators or jaws that can be moved independently. Asymmetric jaws have got many clinical applications in radiation therapy. In the present study, the dosimetric characteristics of asymmetric collimators from our linear accelerator with 6 and 18 MV X-rays were carried out. The field size factors (FSF) and half value layer (HVL) were measured in a water phantom using 0.6 cc Farmer chamber for symmetric and asymmetric fields for both 6 and 18 MV X-rays. Measurements of beam penumbra, percentage depth dose (PDD), cross beam profiles and calculated isodose curves were measured by RFA 300 for both asymmetric and symmetric fields. The FSF were found to agree with in 3% for symmetric and asymmetric fields. The HVL in water was found to be 15.8 cm and 14.4 cm for 6 MV photons and 26 cm and 22.9 cm for 18 MV photons at the central axis and at 20 cm off the central axis. At 30 cm depth the percentage depth dose for symmetric and asymmetric fields were found to differ as high as 6% for 6 MV and 4% for 18 MV fields. No observable difference in penumbra was noticed for symmetric and asymmetric fields of same dimensions. The constrictions of isodose curves at the edge nearer to central axis were noticed for asymmetrically placed fields. The observed differences could be due to the passage of primary beam through differential thickness of the flattening filter which alters the beam quality.
Wydawca

Rocznik
Tom
14
Numer
1
Strony
33-46
Opis fizyczny
Daty
wydano
2008-01-01
online
2008-12-29
Twórcy
  • Department of Radiation Physics, Kidwai Memorial Institute of Oncology, Hosur Road, Bangalore, 560 029, Karnataka, India
autor
  • Department of Radiation Physics, Kidwai Memorial Institute of Oncology, Hosur Road, Bangalore, 560 029, Karnataka, India
autor
  • Department of Radiation Physics, Kidwai Memorial Institute of Oncology, Hosur Road, Bangalore, 560 029, Karnataka, India
autor
  • Department of Radiation Physics, Kidwai Memorial Institute of Oncology, Hosur Road, Bangalore, 560 029, Karnataka, India
autor
  • Department of Radiation Physics, Kidwai Memorial Institute of Oncology, Hosur Road, Bangalore, 560 029, Karnataka, India
autor
  • Department of Radiation Physics, Kidwai Memorial Institute of Oncology, Hosur Road, Bangalore, 560 029, Karnataka, India
Bibliografia
  • Hanson WF, Berkley LW, Peterson M. Off-axis beam quality change in linear accelerator X-ray beams. Med Phys. 1980; 7: 145-150.[PubMed][Crossref]
  • Horton JL. Dosimetry of the Siemens Mevatron 67 linear accelerator. Int J Radiat Oncol Biol Phys. 1983; 9: 1217-1223.
  • Khan FM, Gerbi BJ, Deibel FC. Dosimetry of asymmetric X-ray collimators. Med Phys. 1985; 13: 936-941.
  • Kwa W, Kornelson RW, Harrison, El-khatib E. Dosimetry for asymmetric X-ray fields. Med Phys. 1994; 21: 1599-1604.[Crossref]
  • Loshek DD. Applications in physics of independent collimator features of Varian Clinac 2500. In: Proceedings of 10th Varian users meeting. California: c1984. p. 81-87.
  • Palta JR, Ayyankar KM, Suntharalingam N. Dosimetric characteristics of a 6 MV photon beam from a linear accelerator with asymmetric collimator jaws. Int J Radiat Oncol Biol Phys. 1988; 14: 383-387.
  • Patterson MS, Shragge PC. Characteristics of an 18 MV photon beam from a Therac-20 medical linear accelerator. Med Phys. 1981; 8: 312-318.[Crossref]
Typ dokumentu
Bibliografia
Identyfikatory
Identyfikator YADDA
bwmeta1.element.-psjd-doi-10_2478_v10013-008-0004-2
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.