2010 | 8 | 6 | 1216-1222
Tytuł artykułu

Comparison of IFCC and NGSP methods for determination of glycated haemoglobin using advanced regression techniques

Treść / Zawartość
Warianty tytułu
Języki publikacji
EN
Abstrakty
EN
The objective of this work was a correct statistical comparison of two assay methods for determination of glycated haemoglobin HbA1c. The immunoturbidimetric determination of HbA1c was performed in two ways: using an automatic analyser Hitachi 912, calibrated according to the IFCC reference system (International Federation of Clinical Chemistry and Laboratory Medicine) and using an analyser Advia 1200 and the NGSP reference system (National Glycohaemoglobin Standardization Program). For statistical comparison of these two analytical methods several advanced regression methods were used, which respect random errors of both compared methods. Specifically, Deming regression with and without weights, orthogonal regression, and Passing-Bablok regression were employed. The results demonstrate that the investigated analytical assay methods do not correspond to each other. The summarized results indicate usefulness of better harmonisation of two existing reference systems.
Wydawca

Czasopismo
Rocznik
Tom
8
Numer
6
Strony
1216-1222
Opis fizyczny
Daty
wydano
2010-12-01
online
2010-10-08
Twórcy
  • Department of Chemistry, University of SS. Cyril and Methodius, Trnava, SK-917 01, Slovak Republic
autor
  • Department of Chemistry, University of SS. Cyril and Methodius, Trnava, SK-917 01, Slovak Republic, jan.mocak@ucm.sk
  • Department of Chemistry, University of SS. Cyril and Methodius, Trnava, SK-917 01, Slovak Republic
autor
  • Analytical — diagnostic laboratory, Prešov, SK-080 01, Slovak Republic
Bibliografia
  • [1] D.B. Sacks, Clin. Chem. 49, 1245 (2003) http://dx.doi.org/10.1373/49.8.1245[Crossref]
  • [2] A. Lapolla, P. Traldi, D. Fedele, Clin. Biochem. 38, 103 (2005) http://dx.doi.org/10.1016/j.clinbiochem.2004.09.007[Crossref]
  • [3] M. Penttilä, T. Halonen, K. Punnonen, U. Tiikkainen, Scand. J. Clin. Lab. Invest. 65, 453 (2005) http://dx.doi.org/10.1080/00365510510025809[Crossref]
  • [4] S. Manley, W.G. John, S. Marshall, Diabet. Med. 21, 673 (2004) http://dx.doi.org/10.1111/j.1464-5491.2004.01311.x[Crossref]
  • [5] Consensus Committee: J. Hicks, M. Muller, M. Panteghini, G. John, L. Deeb, J. Buse, D.M. Nathan, R. Kahn, E. Ferrannini, R. Heine, M. Silink, J.C. Mbanya, Diabetes Care 30, 2399 (2007) http://dx.doi.org/10.2337/dc07-9925[Crossref]
  • [6] A. Geistanger, S. Arends, Ch. Berding, T. Hoshino, J.O. Jeppsson, R. Little, C. Siebelder, C. Weykamp, Clin. Chem. 54, 1379 (2008) http://dx.doi.org/10.1373/clinchem.2008.103556[Crossref]
  • [7] J.O. Jeppsson, U. Kobold, J. Barr, A. Finke, W. Hoelzel, T. Hoshino, K. Miedema, A. Mosca, P. Mauri, R. Paroni, L. Thienpont, M. Umemoto, C. Weykamp, Clin. Chem. Lab. Med. 40, 78 (2002) http://dx.doi.org/10.1515/CCLM.2002.016[Crossref]
  • [8] H.G. Schneider, I. Goodall, P.G. Colman, M. McLean, G. Barker, Int. Med. J. 37, 213 (2007) http://dx.doi.org/10.1111/j.1445-5994.2006.01302.x[Crossref]
  • [9] K. Paisooksantivatana, A. Kongsomgan, L. Leohirun, K. Atamasirikul, M. Kunakorn, Diabetes Res. Clin. Pract. 83, 43 (2009) http://dx.doi.org/10.1016/j.diabres.2008.11.028[Crossref]
  • [10] K. Paisooksantivatana, A. Kongsomgan, K. Khupulsup, Diabetes Res. Clin. Pract. 85, 22 (2009) http://dx.doi.org/10.1016/j.diabres.2009.06.005[Crossref]
  • [11] W. Hoelzel, C. Weykamp, J.O. Jeppsson, K. Miedema, J.R. Barr, I. Goodall, T. Hoshino, W.G. John, U. Kobold, R. Little, A. Mosca, P. Mauri, R. Paroni, F. Susanto, I. Takei, L. Thienpont, M. Umemoto, H.M. Wiedmeyer, Clin. Chem. 50, 166 (2004) http://dx.doi.org/10.1373/clinchem.2003.024802[Crossref]
  • [12] Roche Diagnostics GmbH D-M, Cobas Integra 400/700/800: Hemoglobin A1C hemolysate application standardized according to IFCC (Roche Diagnostics Corporation, Indianapolis, IN, USA, 2002)
  • [13] J. Mocak, B. Balla, A. Bobrowski, P. Blazicek, Chem. Pap. 57, 143 (2003)
  • [14] K. Linnet, Clin. Chem. 39, 424 (1993)
  • [15] H. Passing, W. Bablok, J. Clin. Chem. Clin. Biochem. 21, 709 (1983)
  • [16] H. Passing, W. Bablok, J. Clin. Chem. Clin. Biochem. 22, 431 (1984)
  • [17] W. Bablok, H. Passing, R. Bender, B. Schneider, J. Clin. Chem. Clin. Biochem. 26, 783 (1988)
  • [18] J.M. Bland, D.G. Altman, Stat. Methods Med. Res. 8, 135 (1999) http://dx.doi.org/10.1191/096228099673819272[Crossref]
  • [19] Analyse-it for Microsoft Excel ver. 2.12 (Analyse-it Software, Ltd., United Kingdom, 1997–2010) http://www.analyse-it.com
  • [20] Medcalc ver. 10.4.0.0. (MedCalc Software bvba, Mariakerke, Belgium, 2005–2010) http://medcalc.be
  • [21] JMP 8 (SAS Institute Inc., Cary, NC, USA, 2005) http://jmp.com
  • [22] K. Linnet, Clin. Chem. 45, 882 (1999)
  • [23] G.S. Dhatt, M.M. Agarwal, B. Bishawi, Clin. Chim. Acta 358, 81 (2005) http://dx.doi.org/10.1016/j.cccn.2005.02.009[Crossref]
  • [24] A. Adamikova, J. Bakala, J. Bernatek, J. Rybka, S. Svacina, Ann. Nuc. Med. 20, 615 (2006) http://dx.doi.org/10.1007/BF02984659[Crossref]
  • [25] J.M. Bland, D.G. Altman, Lancet 307 (1986) [Crossref]
  • [26] D. Stöckl, K. Dewitte, L.M. Thienpont, Clin. Chem. 44, 2340 (1998)
Typ dokumentu
Bibliografia
Identyfikatory
Identyfikator YADDA
bwmeta1.element.-psjd-doi-10_2478_s11532-010-0096-x
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.